Mc-techgroup CLF-5_W Data sheet Bedienungsanleitung

Stöbern Sie online oder laden Sie Bedienungsanleitung nach Ausrüstung Mc-techgroup CLF-5_W Data sheet herunter. M&C TechGroup CLF-5_W Data sheet User Manual Benutzerhandbuch

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 2
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 0
M&C TechGroup Germany GmbH • Rehhecke 79 • 40885 Ratingen • Germany
info@mc-techgroup.com • www.mc-techgroup.com • Fon +49 2102 935-0 • Fax +49 2102 935-111
7.7
Embracing Challenge
High retention rate of 99,9999% for
particles > 0,1µm
Also with integrated hydrophobic dia-
phragm for analyser protection
Condition of lter element visible from
outside
Easy change of lter element
Wall mounting
Special Features
Fluid Particle Filter Series CLF®
Version CLF-5, CLF-5/W for removal of aerosols from gases
CLF-5
Application
The M&C fluid particle filter CLF-5, CLF-5/W
is suitable for filtration of fluid particles of
all types and is recommended for sample
gases with an acid dew point above 100 °C.
Examples are measurements in flue gas of
heavy oil and black coal combustions.
The filter separates the aerosols (very fine
fluid particles) which still pass the gas cooler.
The most effective position of the CLF-.. fil-
ter is downstream the sample conditioning
close to the flowmeter of the analyser. For
additional system protection we provide the
version CLF-5/W with integrated hydrophobic
diaphragm, working as a liquid stop.
Description
The filter element of the M&C fluid particle
filter is constructed in two sections with a
flow direction from the inside to the outside
of the filter element. The inner, very fine, fibre
layer binds the fluid particles suspended in
the gas and leads them together with the gas
flow to the outer, larger layer. On their way
through the filter element, the very fine fluid
particles accumulate with others and form
droplets. The vertical flow direction and the
force of gravity cause the droplets to drip into
the filter pot.
The filter element remains fully effec-
tive even when completely saturated
with fluid. If it is not effected by solid par-
ticles, the lifetime is nearly unlimited.
The compressed Micro-Fibres are made with
binding of PVDF in order to prevent influ-
ences on the sample gas.
The version CLF-5/W is equipped with an
integrated liquid stop for water and water
identical liquids. The modified filter ele-
ment clamp has a protective hydropho-
bic diaphragm. In case the sample condi-
tioning system does not work proper, the
CLF-5/W will stop the liquid in front of the
filter outlet.
The condition of the filter is visible through
the glass body without opening the filter.
The separated acid mist can continuously be
discharged with an external mounted peristal-
tic pump SR25.1 (option) connected by a GL25
adapter. No tools are required to change the
filter element. The optimised position of the
O-ring always guarantees a safe sealing of
the filter body to the filter head. The filter
in- and outlet can be turned about 180 °
on the mounting bracket to achieve easy
mounting and flexible adaptation to local
circumstances.
Handle with care! Danger of acid burns!
Technical specications and illustrations are without
obligation, subject to modications. 09.96/06.06
Seitenansicht 0
1 2

Inhaltsverzeichnis

Seite 1 - Description

M&C TechGroup Germany GmbH • Rehhecke 79 • 40885 Ratingen • Germany [email protected] • www.mc-techgroup.com • Fon +49 2102 935-0 • Fax +49 21

Seite 2 - Technical Data

M&C TechGroup Germany GmbH • Rehhecke 79 • 40885 Ratingen • Germany [email protected] • www.mc-techgroup.com • Fon +49 2102 935-0 • Fax +49 21

Kommentare zu diesen Handbüchern

Keine Kommentare