9-3.12-ME M&C Products Analysentechnik GmbH- 1 -INSTRUCTION MANUALMicroprocessor controlled oxygen analyser PMA®Version PMA100 (V1.52)
M&C Products Analysentechnik GmbH 9-3.12-ME - 10 -8.2.1 mA outputThe mA output is available on the back panel of thePMA100 (see fig.8) at the 15-p
9-3.12-ME M&C Products Analysentechnik GmbH- 11 -9. Receipt and storageThe PMA100 is completely pre-installed and normally de-livered in one packa
M&C Products Analysentechnik GmbH 9-3.12-ME - 12 -12.Menu description12.1 Table of content• Overview menu-drive and Overview operating keys ...
9-3.12-ME M&C Products Analysentechnik GmbH- 13 -Overview menu-drive Overview operating keys• Startup of measuring menu• Function: to leave the ac
M&C Products Analysentechnik GmbH 9-3.12-ME - 14 -12.2 Warming upAfter switching on the analyser type PMA100, the warming-up begins.The following
9-3.12-ME M&C Products Analysentechnik GmbH- 15 -SelectSelect12.4 Parameterising level 112.4.1 LanguageThe following languages are selectable by t
M&C Products Analysentechnik GmbH 9-3.12-ME - 16 -relay config.relay 1A selection is done by the Direction keys ⇑⇓ and confirmedby the Enter key.T
9-3.12-ME M&C Products Analysentechnik GmbH- 17 -↑↑input config.input 1...4↑↑averageno, 1, ..., 100 secondsAn alarm situation can be kept. If yes,
M&C Products Analysentechnik GmbH 9-3.12-ME - 18 -conc. alarm 1...4[< or >] xx.xx%A jump to the next position happens via the Select key.The
9-3.12-ME M&C Products Analysentechnik GmbH- 19 -12.5 Parameterising level 212.5.1 Step into the menuSystem parameters can be changed in level 2.
M&C Products Analysentechnik GmbH 9-3.12-ME - 2 -This instruction manual does not claim completeness and is subject to technical modifications.©5/
M&C Products Analysentechnik GmbH 9-3.12-ME - 20 -12.5.5 Current signal with calibrationDuring the calibration a mA-signal can be put out as:off m
9-3.12-ME M&C Products Analysentechnik GmbH- 21 -• Current pressureThe Enter key opens the display to change the value forthe current pressure. Th
M&C Products Analysentechnik GmbH 9-3.12-ME - 22 -12.7.3 STX and ETX characterIt is necessary to determine a start respectively stop signalfor the
9-3.12-ME M&C Products Analysentechnik GmbH- 23 -12.8.3 Device modeThis menu point selects via the Direction keys ⇑⇓ whetherthe analyser works as
M&C Products Analysentechnik GmbH 9-3.12-ME - 24 -zero conc.xx.xx%O2Calzero conc.(-)xx.xx%O2EnterSelect Enter⇑⇑⇑⇑⇑⇓⇓⇓⇓⇓Selectspan-conc.00.00% O2En
9-3.12-ME M&C Products Analysentechnik GmbH- 25 -• Manual calibration offsetEnter starts the calibration procedure.After reaching a stable level f
M&C Products Analysentechnik GmbH 9-3.12-ME - 26 -cal. log-bookM 10.11.99 11:o7Selectcalib. diff.ofs: (-)x.x% g: x.x%Selectinflate gas . . . sec.2
9-3.12-ME M&C Products Analysentechnik GmbH- 27 -13.Calibration13.1 CalibrationFour calibration modes are available with the analyser PMA100:• man
M&C Products Analysentechnik GmbH 9-3.12-ME - 28 -Gas Formula + 20°C+50°CArgon Ar - 0,23 - 0,25Acetylene C2H2- 0,26 - 0,28Acetone C3H6O - 0,63 -
9-3.12-ME M&C Products Analysentechnik GmbH- 29 -Example :The residual oxygen percentage should be measured in aclosed 100% carbon dioxide (CO2) a
9-3.12-ME M&C Products Analysentechnik GmbH- 3 -1. Electrical standardsThe electrical standard corresponds to the safety regula–tions concerning t
M&C Products Analysentechnik GmbH 9-3.12-ME - 30 -14.MeasuringThe connections for sample gas inlet and outlet are locatedon the back panel of the
9-3.12-ME M&C Products Analysentechnik GmbH- 31 -15.Function of in- and output contacts andalarmsThe following in- and output contacts are availab
M&C Products Analysentechnik GmbH 9-3.12-ME - 32 -The following alarm messages are available at the commonalarm output:• concentration alarm 1 - 4
9-3.12-ME M&C Products Analysentechnik GmbH- 33 -17.Maintenance and repairBefore carrying out any maintenance or service activities,the engineer c
M&C Products Analysentechnik GmbH 9-3.12-ME - 34 -18.Trouble shootingFault description Display SolutionMeasuring cell doesn‘t work E:01 resp. F:01
9-3.12-ME M&C Products Analysentechnik GmbH- 35 -19.Spare part listWear, tear and replacement part requirements depend onspecific operating condit
M&C Products Analysentechnik GmbH 9-3.12-ME - 4 -4. Used terms and signal indicationThese are persons with necessary qualification, who arefamilia
9-3.12-ME M&C Products Analysentechnik GmbH- 5 -6. ApplicationThe transducer of the PMA100 works at a stable tempe-rature of +55°C. Therefore the
M&C Products Analysentechnik GmbH 9-3.12-ME - 6 -the corresponding output of the amplifier, which is a currentand also proportional to the oxygen
9-3.12-ME M&C Products Analysentechnik GmbH- 7 -7.3 Dimensions and weightThe analyser is build in a 19” housing, also suitable fortable mounting.
M&C Products Analysentechnik GmbH 9-3.12-ME - 8 -Part No. 03A3000(a): PMA100, power supply 230VAC, 50Hz, 115VAC, 60Hzsignal: 4-20mA; (a)=115VMeasu
9-3.12-ME M&C Products Analysentechnik GmbH- 9 -8. Supply connections8.1 MediumNote!The oxygen analyser PMA100 is suitable for continuousmeasureme