14.1a ME Gas sampling and gas conditioning technology 37
16.7 SETTING SERVICE INTERVAL, LANGUAGE, CONTROL TEMPERATURES AND
TEMPERATURE SENSOR TYPES
To change service interval, language, control temperatures and temperature sensor types:
Disconnect the device from the mains.
Hold the PRG key down and switch the device back on until the message to release the
key is displayed. The following is displayed (2nd screen ):
Select the parameter to be changed using the arrow key and confirm using the PRG
key.
Use the arrow keys, the settings and values can be changed and confirmed with the
PRG key . Press and hold down the PRG key for 5 seconds to apply all changes.
To discard the changed settings, keep the ESC key pressed for 5 seconds